HP8648 is an entry level signal generator with a good price on Ebay. My first HP8648C (9KHz-3,2GHz) bought in good condition was a pleasure, easy to operate, good accuray and enough frequency range. It can be remote controled via GPIB (HP-IB) bus. After this experience i found one more HP8648B (9KHz-2GHz), very cheap, but "no output" according seller description. I didn't notice other error in the seller's photos, maybe only the stepped attenuator has an issue. In reality was a "small" disaster. Below is a summary of my experience during HP8648B repairing.
#1. Output stepped attenuator out of order, as expected. Output level has abnormal attenuation steps.
#2. Error 627 during power up due to missing battery
#3. Error "519 Output LF level cal restore failure"
#4. Output level can't be adjusted by vernier if the frequency is less than 1GHz.
#5. Error "608 Synth out-of-lock: Main"
#6. Error "605 ALCout-of-lock: Div, HET, Main, X"
#7. Error "518 Atten comp loss cal restore failure" (20.11.2018)
As you can know for HP8648A/B/C/D and HP8647, service manual has not included schematic diagrams and general content is poor in case you need to repair something. HP advice is to replace faulty module, very simple... Also it is almost impossible to calibrate without dedicated instruments. HP8648 is designed for automated calibration via HP-IB bus. Indeed, very bad news for us but not impossible. We will see how to fix above malfunction, without to be expensive.
Some modules used by Agilent 8648 are similar with Agilent 8920, but not exactly the same. Also Agilent 8920A Service Manual is better than 8648.
For example here are diagrams for Reference Unit, Signal Generator Synth and Output RF extracted from 8920 manual. See also the 8920 schematics!
![]()
HP8648 Operating and Service Guide
Agilent 8920A Service Manual - schematics included!
Agilent 8920A manual - block diagrams
10MHz Timebase adjust. Good quality VHF frequency counter required.
#1. Output attenuator marked as "Defect". Part No. 33322-60011
Resistive pads (10/20/40/60dB) were found in good order, no burned or cracked resistors. Abnormal behaviour observed when 20dB step is engaged. As you can expect i found bad rubber O-rings and solenoid coil didn't switched one contact. Path of signal disturbed by this wrong open contact.
Solution: open attenuator assembly and replace all small Orings (OD 2mm / ID 1mm/ CS 0.5mm).
O-rings source: AliExpress seller (01.09.2018, dead link), 2x1x0.5mm NBR, 50pieces, ~10$
Internal view of attenuator:
![]()
Defect Stepped Attenuator
Bad contact - not fully engaged
How to take out magnets and driver board
Missing O-rings, commun problem of HP stepped attenuators
Better view of rubber orings, all pieces are in poor condition
#2. Error 627 during power up due to missing RAM memory battery
Battery is not in place, somebody try to replace it. Also i found free and losen bolts inside of generator box. We will see below what happen if you power up motherboard and few bolts are free to move under PCB.
During power up the software try to reload calibration parameters from every module Motherboard/Reference/Synthesizer/Output/Extension. Every module has unique "ROM". Battery is used to keep SRAM memory alive, user settings and configuration of installed modules. Without battery RAM memory will be reloaded during start-up using default "ROM" instructions.
BR-2/3A 3V Lithium. Do not try to recharge! Replace battery and will be fine.
#3. Error "519 Output LF level cal restore failure"
This error is a real problem, it is not even mentioned in the manual! Searching over Internet i found only one answer: "A6 Output Module has calibration problems or motherboard need to be replaced"... I can't run any calibration via HP-IB bus and to replace motherboard is to much for me. My HP8648B will be available for spare parts?! :). I didn't notice any anomaly of generator if 519 error is active, i'm not sure 100% because i didn't check 9KHz-100KHz output level.
Don't worry, i have one more generator HP8648C, so let's start to experiment. One by one i start to replaced module between generators in order to locate 519 Error. If I replace "A6 Output Assembly", the error disappears! Very good, i know were to dig deep. What is inside of A6 Output Module, how calibration are stored? ROM memory is used (according service manual) to store calibration parameters. In reality ROM memory is a EEPROM microwire memory 93C66 or 93C46 . Hmm i have some ideas to copy calibration file from good A6. I need to mention, if you take out EEPROM and you try to start A6 module, same 519 will be received.
![]()
A6 Output Module. Rusted screen shields but PCB in good condition and requires a good cleaning. 93C66 location visible.
HP8920 Output Module, similar with A6/HP8648, schematic diagram (missing ALC INT/EXT switch DG413, anyway A6 INT ALC is not used for B/C/D, only for A version)
For 93Cx6 programming I use an old laptop with native serial port (no USB/RS232 adapter), Windows XP and PonyProg 2000 software. Choose 16 bit mode of 93Cxx leave pin no.6 OPEN (floating) or to VDD, do not connect to GND (8bit mode).
Any EEPROM 93Cxx programmer can be used.
![]()
Schematic of 93Cxx programming interface
PCB Sprint-Layout 6 file
PonyProg 16 bit selection
After EEPROM programming, 519 error is clear, well done! OK is not perfect solution to use calibration file from other module. Testing between 9KHz-100KHz reveals poor calibration of output level under 25KHz. Corrupted A6 Module type 08648-60116.
-NEW UPDATE (11.09.2018) I have one more HP8648C (LCD version) in good condition. LF response of defective HP8648B is more better using that calibration file.
![]()
A6 Output Assembly 08648-60192 (without variable resistors), EEPROM 93C66 "bin" dump
NEW A6 Output Assembly 0648-60131 (with variable resistors), EEPROM dump
Inside of HP8648 you can locate 6 pieces 93Cxx EEPROMs, connected to SCX6206TQM IO Chip. I found one more version of A5 module using AMI IO Chip. Because is a big trouble to loose one of memory parameter, i made full backup of all 12 pieces EEPROM, from both generators HP8648B and HP8648C. Original corrupted file of A6 HP8648B is not available. Below you can download my BIN files in case you have trouble. Please take care and don't blame me if something bad happens! Before any EEPROM programming, make a backup of original file, even if is corupted or unknown condition! Also i recommend to test programming interface using spare 93CX6.
You can mix all modules between HP8648 B/C, only A10 Frequency Extension will not work above 1GHz. Also if you take out A10, RF signal is still active on SMC connector (from A6 output).
![]()
LE: 23.08.2025 - programming by XGECU TL866+ required selection of "8bit" and "SWAP". Different way to dump EEPROM betewwn PonyProg and XGECU?
Anyway, try different settings 8/16, SWAP, util the EEPROM is correct recognized by 8648.
| EEPROM CALIBRATION BIN DUMPS FROM HP8648B/C (VFD version) | |||
|---|---|---|---|
| HP8648B SN: 3426A00391 |
Observation | HP8648C SN: 3847U3029 |
Observation |
| Motherboard U59 | Located under Attenuator MB revision B.04.09 |
Motherboard U59 | Located under Attenuator MB revision B.04.09 |
| Motherboard U201 | Located under Fan MB revision B.04.09 |
Motherboard U201 | Located under Fan MB revision B.04.09 |
| A4 REFERENCE | Option 1E5 08648-60042 93C46 |
A4 REFERENCE | STANDARD 08647-61045 93C46 |
| A5 SIG GEN SYNTH | 08648-60115 SCX6206TQM IO Chip |
A5 SIG GEN SYNTH | 08648-60115 AMI IO Version |
| A6 OUTPUT | Not Available/Corrupted 08648-60116 (with variable resistors) |
A6 OUTPUT | 08648-60192 (without variable resistors) |
| A10 FREQ. EXTENSION | 08647-60213 | A10 FREQ. EXTENSION | 08648-63191 |
|
EEPROM CALIBRATION BIN DUMPS - NEW HP8648C (LCD version) Not full backup for the moment (11.09.2018) |
|
|---|---|
| HP8648C SN: 3462U01161 |
Observation |
| Motherboard U59 | Located under Attenuator MB revision B.04.01 |
| Motherboard U201 | Located under Fan MB revision B.04.01 |
| A4 REFERENCE | 08647-61045 |
| A5 SIG GEN SYNTH | 08648-60115 |
| A6 OUTPUT | 08648-60131 (with variable resistors) |
| A10 FREQ. EXTENSION | 08648-60024 |
#4. Output level can't be ajusted by vernier if the frequency is under 1GHz
Level adjutement of output is controled by 0-110dB stepped attenuator and ALC circuit between 10dB steps or above -6dBm vernier value. The story is very long and i will give only the solution and some tips.
Above 1GHz, ALC circuit is included inside of A10 Frequency Extension Module. Below 1GHz, ALC signal from output of A10 is sent back (feedback) to A6 in order to be compared with DAC value. Output of A6 can be checked on SMC connector, under A10 input. There you can find:
-
fixed amplitude value if output is >1GHz or A10 module missing/faulty under 1GHz
-
10dB range adjustable leve (0.1dB steps) if frequency is <1GHz. Note: if vernier value is between -6dBm and maximum output value, level is controled by ALC, stepped attenuator is bypassed.
After many measurements, on pin 5 J29 i found 0 Volts ALC signal during operation below 1GHz. I took out A10 and 0...3V voltage was applied to simulate ALC feedback. A6 Output level starts to jump and that confirmed good operation of A6.
Malfunction was located to A10. Here i found +8V power supply missing from pin 7,8 J29. Voltage regulator U203 (located under FAN, difficult to be observed) is supplied from +15V bus through 3,3ohm power resistor (located under PSU). My resistor was blown up, i guess by a heavy short circuit during seller testing with free bolts under motherboard!
Resistor was replaced and ALC 0.1dB steps are working again below/above 1GHz.
![]()
3R3 Resistor Burned
Replaced Resistor
A10 Frequency Extension Modules - internal view
#5. Error "608 Synth out-of-lock: Main"
Error occurs one day after above issues...Malfunction located inside of "A5" main VCO 500-1000MHz. Also red LED OOL (out of lock) has steady light if VCO frequency is between 850-1000MHz. In this range tuning voltage was not enough, can't reach more than 15V. Normal value of VCO varicap voltage is aprox. 3.5...20V.
Again after many hours of investigation i found a strange and rare defect: current leakage of one varicap diode!
A5 VCO is located inside of aluminum box, here are used 8 varicap diodes in order to achieve 0.5-1GHz frequency span. Tuning voltage was pulled down by one of diodes, but only if control voltage was more than 12V. I tried to replace the defective varicap (BB809) with other available diodes, but the output level started to drop during the frequency increasing and OOL appears. I found a BB535 SMD diode from an old Motorola mobile phone. The result is very good.
LE: 25.08.2025 New diode BB809 installed.
![]()
Testing procedure of varicap diodes for current leakage. 100nH inductance disconnected in order to check voltage drop across 5K6 resistor installed between 25V PSU and tuning diodes. Voltage drop must to be 0V.
New BB535 (S white marking) diode installed
Schematic of voltage tunning circuit (reverse engineering), voltage control values and VCO OFFSET adjustment.
HP8920A Synth Board schematic - similar with HP8648 A5 module
#6. Error "605 ALCout-of-lock: Div, HET, Main, X"
Error occurs during VCO out of lock malfunction. Was solved by remove/install A10 unit. Under A10 module i found an improper screw, the length was too long and connectors between A10 and motherboard didn't have a good contact.
Check A10 Frequency Extender position or future investigations are required. Check also motherboard jumper J30 to be CLOSE.
#7. Error "518 Atten comp loss cal restore failure"
My generator has not been used for several months. Last time everything was fine. I recently needed to start it and I had a new surprise. Error "518" attenuator compensation calibration parameters can't be restored during start-up.
The error can be solved if EEPROM 93C66 (U201) located on the motherboard is rewritten using a backup dump.
Many problems with EEPROM corupted contents and I don't know why...Again, my advice is to open the generator and to backup all 93Cxx EEPROM.
![]()
HP8648 Test Package v7.01 / EPSG090522 (passwords included), useful for calibration
HP8648 software test utility - WinXP screenshots
GPIB/HPIB address info (HP-Agilent groups.io)
Calibration Demo (mp4 file, unknown source)
HP8648 ROM Board Dump B04.09 (HP 08647-60117 / A3A1).
yo4hfu@2010-2025