Keithley Instruments - 2000 6-1/2 Digit Multimeter
2000 6 1/2-Digit Multimeter

The Model 2000 61/2-digit Multimeter is part of Keithley’s family of high performance DMMs. The 2000 is a fast, accurate, and highly stable instrument that’s as easy to operate as it is to afford. It combines broad measurement ranges with superior accuracy specifications — DC voltage from 100nV to 1kV (with 0.002% 90-day basic accuracy) and DC ohms from 100µohm to 100Mohm (with 0.008% 90-day basic accuracy).

High Throughput
The 2000 offers exceptional measurement speed at any resolution. At 61/2 digits, it delivers 50 triggered rdgs/s over the IEEE-488 bus. At 41/2 digits, it can read up to 2000 rdgs/s into its internal 1024 reading buffer.

Creating a self-contained multipoint measurement solution is as simple as plugging a scanner card into the option slot on the 2000’s back panel. This approach eliminates the complexities of triggering, timing and processing issues and helps reduce test time significantly.

High Measurement Flexibility
The 2000 has 13 built-in measurement functions, including DCV, ACV, DCI, ACI, 2Wohm, 4Wohm, temperature, frequency, period, dB, dBm, continuity measurement, and diode testing.

The frequency counter function can make accurate signal measurements up to 300 kHz, eliminating the need for a separate counter.

High Compatibility
TestPoint™ Instrument Driver Libraries and run-time programs are included with the 2000 to simplify IEEE-488.2 and RS-232 program generation. A variety of ready-to-run applications programs are also included. A LabVIEW library for the 2000 is available.

High Reliability
The 2000 conforms to rigorous standards governing safety and reliability, including UL-1244, CSA, IEC-1010, FCC and VDE. A 1000V input rating guards against accidental overloads, and the instrument recovers from overloads in milliseconds. It can accommodate 1500V from high-to-earth for added protection and increased reliability.

 

DC VOLTAGE Accuracy: ±(ppm of reading + ppm of range), 1 PLC
Range Resolution Input Resistance 1 Year 23° C ± 5
100.0000 mV 0.1 µV > 10 Gohm 50 + 35
1.000000 V 1.0 µV > 10 Gohm 30 + 7
10.00000 V 10 µV > 10 Gohm 30 + 5
100.0000 V 100 µV 10 Mohm 35 + 6
1000.000 V 1 mV 10 Mohm 45 + 10

RESISTANCE Accuracy: ±(ppm of reading + ppm of range)
Range Resolution Test Current 1 Year 23°C ± 5
100.0000 ohm 100 µohm 1 mA 100 + 40
1.000000 kohm 1 mohm 1 mA 100 + 10
10.00000 kohm 10 mohm 100 µA 100 + 10
100.0000 kohm 100 mohm 10 µA 100 + 10
1.000000 Mohm 1 ohm 10 µA 100 + 10
10.0000 Mohm 10 ohm 700 nA 400 + 10
100.000 Mohm 100 ohm 700 nA 1500 + 30

CONTINUITY ±(ppm of reading + ppm of range)
Range Test Current 1 Year 23°C ± 5
1 kohm 1 mA 120 + 100

DC CURRENT Accuracy: ±(ppm of reading + ppm of range)
Range Resolution Burden Voltage 1 Year 23°C ± 5
10 mA 10 nA < 0.15 V 500 + 40
100 mA 100 nA < 0.03 V 500 + 50
1 A 1 µA < 0.3 V 800 + 40
3 A 10 µA < 1 V 1200 + 15

DC Speed and Noise Rejection
Int. Rate Max. Rdgs./s Digits RMS Noise @ 10V NMRR CMRR
10 PLC 5 61/2 2 µV 60 dB 140 dB
1 PLC 50 61/2 6 µV 60 dB 140 dB
0.04 PLC 1000 51/2 30 µV 80 dB
0.01 PLC 2000 41/2 1 mV 80 dB

TRMS AC VOLTAGE Accuracy: ± (% of reading + % of range)
Range Resolution 3Hz–10Hz 10Hz–20kHz 20kHz–50kHz 50kHz–100kHz 100kHz–300kHz
100 mV 0.1 µV
1 V 1 µV
10 V 10 µV 1 Year, 23°C ±5°C (all ranges)
100 V 100 µV 0.35+0.03 0.06+0.03 0.12+0.05 0.6+0.08 4+0.5
750 V 1 mV
Max. 5:1 crest factor at full scale for inputs >5% of range.

TRMS AC CURRENT Accuracy: ±(% of reading + % of range), 1 Year
Range Resolution 3 Hz–10 Hz 10 Hz–5 kHz
1.000000 A 1 µA 0.30 + 0.04 0.10 + 0.04
3.00000 A 10 µA 0.35 + 0.06 0.15 + 0.06

DIODE TEST ±(ppm of reading + ppm of range)
Range Test Current 1 Year 23°C ± 5
3 V 1 mA 40 + 7
10 V 100µA - 10µA 40 + 7

FREQUENCY AND PERIOD
ACV Range Frequency Range Period
Range
Gate
Time
Resolution
±(ppm of rdg.)
Accuracy
(1 Year)±(% of rdg.)
100 mV 3 Hz 333 ms
to to to 1 s 0.3 0.01
750V 500 kHz 2 µs

Temperature Characteristics Thermocouple 2, 3, 4 90 Day/1 Year (23°C ± 5°C)
      Accuracy
Type Range Resolution Relative to
Simulated
Reference Junction
Using
2001-TCSCAN
J –200 to + 760°C 0.001°C ±0.5°C ±0.65°C
N –200 to + 1300°C 0.001°C ±0.5°C ±0.70°C
T –200 to + 400°C 0.001°C ±0.5°C ±0.68°C

Maximum System Speeds
Range Change: 50/s.
Function Change: 40/s.
Autorange Time: <30ms.
ASCII readings to RS-232: 55/s.


Specification Highlights:

MATH FUNCTIONS: Null, Limit, and Min/Max/Avg/Std Dev of stored readings.

DATA STORAGE: Selectable up to 1024 readings max.

STANDARD PROGRAMMING LANGUAGES: SCPI, Fluke 8840A/8842A, Keithley 199/196.

INTERFACES: IEEE-488 and RS-232.
Power Supply: 100V / 120V / 220V / 240V.
Line Frequency: 45Hz to 66Hz and 360Hz to 400Hz.
Warranty: 3 years.

PHYSICAL
Case Dimensions: 89mm high x 213mm wide x 370mm deep (31/2 in x 83/8 in x 149/16 in).
Unit Weight: 2.9kg (6 lbs 3 oz).

STANDARDS:

Safety: Designed to CSA, UL-1244, IEC-1010.
EMI: Conforms to VDE 0871B (per Vfg 1046/1984). Meets FCC part 15 class B.
          Conforms to CISPR22, EN-5504.
ESD: Conforms to IEC-801-2.

Specifications are subject to change without notice.


Complete product information and specifications on this product is available on-line via the following PDF link. To access it, you must first download and install the appropriate Adobe Acrobat Reader for your operating system. This reader is available free at http://www.adobe.com/prodindex/acrobat/readstep.html

 